| To: | Dan Sawyer <dansawyer@earthlink.net>, amps <amps@contesting.com> |
|---|---|
| Subject: | Re: [Amps] help diagnosing a bad 8450a probe ? |
| From: | Peter Chadwick <g3rzp@g3rzp.wanadoo.co.uk> |
| Reply-to: | g3rzp@g3rzp.wanadoo.co.uk |
| Date: | Fri, 9 Dec 2005 13:34:13 +0100 (CET) |
| List-post: | <mailto:amps@contesting.com> |
Dan said: > However one shows a 1V peak to peak distorted sine like wave. This probe actually works, however due to the fault it has a phase frequency error< Have you checked the four Schottky diodes in the sampler? I've had funny answers like yours when one of them had died. BTW, you need to replace all four with a matched quad, not change just one. 73 Peter G3RZP _______________________________________________ Amps mailing list Amps@contesting.com http://lists.contesting.com/mailman/listinfo/amps |
| <Prev in Thread] | Current Thread | [Next in Thread> |
|---|---|---|
| ||
| Previous by Date: | Re: [Amps] help diagnosing a bad 8450a probe ?, Steve Thompson |
|---|---|
| Next by Date: | Re: [Amps] help diagnosing a bad 8450a probe ?, Dan Sawyer |
| Previous by Thread: | Re: [Amps] help diagnosing a bad 8450a probe ?, Alex Eban |
| Next by Thread: | Re: [Amps] help diagnosing a bad 8450a probe ?, Dan Sawyer |
| Indexes: | [Date] [Thread] [Top] [All Lists] |